Voluntary Worldwide Field Corrective Action of the Vital Signs Hygroscopic Condenser Humidifier (HCH) / Anesthesia Breathing Circuit

TOTOWA, N.J.--()--In April 2011, Vital Signs Devices, a GE Healthcare Company, initiated a voluntary worldwide field correction of the Vital Signs Hygroscopic Condenser Humidifier (HCH)/Anesthesia Breathing Circuit. This device is sold under various part numbers - see list below.

The Vital Signs Devices Passive Humidification Device (Hygroscopic Condenser Humidifier or “HCH”) may have an occlusion that can prevent proper flow of the medical gases or oxygen, possibly resulting in insufficient oxygen delivered to the patient (hypoxia or hypoxemia). This device may be included in Vital Signs anesthesia circuits or sold separately. The affected units were manufactured between January 2011 and April 2011.

GE Healthcare initiated the field correction in April 2011, and began notifying customers with affected units through an Urgent Medical Device Correction Letter and a follow-up call to confirm receipt of the letter. Through these efforts, all customers were informed of the issue and provided with safety instructions. Customers have been directed to NOT use the anesthesia circuits containing the Vital Signs Devices Passive Humidification Device (Hygroscopic Condenser Humidifier, or “HCH”), or the stand alone HCH device with the product item numbers and lot numbers listed below, isolate all affected product, contact Vital Signs Customer Service at 1-800-932-0760 to arrange for the return and replacement of product. All affected Hygroscopic Condenser Humidifier and Anesthesia Breathing Circuits with affected Hygroscopic Condenser Humidifiers are being replaced.

The U.S. Food and Drug Administration (FDA) has classified this recall as a Class I recall. FDA defines Class I recalls as “a situation in which there is a reasonable probability that the use of or exposure to the volatile product will cause serious adverse health consequences or death.”

For additional information regarding this field correction, please contact Vital Signs Customer Service at 1-800-932-0760. Hours of Operation: 8:00 am EDT to 5:00 pm EDT, Monday through Friday.

Product Lot #       Product Lot #       Product Lot #       Product Lot #

5020AEHM 

11087A

A5632X15 

11070A

A5U5201XX1 

11076B

ADU12914 

11074A
11098B 11095A 11083A 11074B
5701 5301C A565192X 11094A 11088A 11075A
5344E 11095A 11090A 11087A
5348V 11095B 11101A 11090B
5353K 11096A A5W32014 11077A 11094B
5366T 11097A 11090B 11098B
5701E 5346Z A5U32020 11070A A5Z51914 11069B 11101A
5702 5301D 11070B 11070A 11101B
A0F52014 11091A 11074B 11084A ADU52914 11073A
11098B 11077B 11097B 11074A
A4112X14 11070A 11081A A8T52914 11082B 11075A
A415112A 11098B 11081B A8U52015 11069A 11077A
A41X2X24 11092A 11083A 11075A ADW32014 11073A
A4612X2C 11077A 11087A 11075B 11083A
11081B 11087B 11083B 11087A
11095B 11090A 11091A 11094A
11098A 11092A 11101B 11098A
A4F12014 11075A 11094B A8U5211X 11094A 11098B
11084B 11097A A8U52X14 11075A 11101A
11090B 11098A 11077A AFN52024 11074A
11091A 11101A 11083A 11083B
11094A 11102A 11090B 11084A
11101B 11102B A8W52914 11069A 11091B
11091A A5U32414 11077B 11083A 11096A
11101A 11080A 11088A 11098B
A4F52X10 11075B 11082A 11094A AFN5291C 11069B
11077B 11082B A9U5211X 11080A 11070B
11082A 11090A 11090B 11083A
11090A 11094A 11094A 11090A
A4F52X14 11074B 11094B AD132X14 11073A 11098A
A4U12214 11080B 11097A 11074A AFN52X1X 11070B
11092A 11097B 11075A 11088A
11097A 11098A 11088A AFR5191X 11069B
11098A A5U32X14 11076B 11095B AFR58014 11069B
11101A 11083B 11096B 11076A
A4U52X1X 11083A 11084A AD152914 11074B 11098B
11087A A5U32X24 11070A AD632X11 11070A 11101A
11090A 11077A 11073A 11101B
A5152914 11077B 11080A 11074A AFR5XX20 11073B
11080A 11083B 11081A 11101A
A52X2314 11081A 11087B 11083A AGN1QX1C 11070A
11091A 11098B ADA52014 11076A 11070B
11096B 11101A ADG5291X 11101B
 

Contacts

GE Healthcare
Annette Busateri, +1 262-442-0966
annette.busateri@ge.com

Contacts

GE Healthcare
Annette Busateri, +1 262-442-0966
annette.busateri@ge.com